Features ABS IP65 Housing and a 5-digit LCD display. Passive transmitter with 18/30V DC and 4-20mA Output. Comes with probe remote.
Features ABS IP65 Housing and a 5-digit LCD display. Passive transmitter with 18/30V DC and 4-20mA Output. Comes with probe remote.
Features
Applications
Ideal for the following applications: Power Plants, Chemical PetroChemical, Pharmaceutical, Biotechnology, Cement, Pulp & Paper, Machine building, Food and Beverage
Temperature Specifications | |
Measuring Range | -4 to 176°F (-20 to 80°C) |
Accuracy | ±0.5°F (±0.3°C) from -40 to 158°F (-40 to 70°C)
±1°F (±0.5°C) beyond |
Unit of Measurement | °C / °F |
Response Time | 1/e (63%) 15 seconds |
Sensitive Element | NTC |
Resolution | 0.2°F (0.1°C) |
Type of Fluid | Air and neutral gases |
Humidity Specifications | |
Measuring Range | From 5 to 95% RH |
Analogue Output | From 0 to 100% RH |
Accuracy | ±1.5% RH if 59 ≤ T ≤ 77°F (if 15 ≤ T ≤ 25°C) |
Drift Linked to Temperature | ±0.04 x (T-20) % RH if T < 59 or T > 77°F (if T < 15 or T > 25°C) |
Unit of Measurement | % RH |
Temperature Specifications | |
Measuring Range | -4 to 176°F (-20 to 80°C) |
Accuracy | ±0.5°F (±0.3°C) from -40 to 158°F (-40 to 70°C)
±1°F (±0.5°C) beyond |
Unit of Measurement | °C / °F |
Response Time | 1/e (63%) 15 seconds |
Sensitive Element | NTC |
Resolution | 0.2°F (0.1°C) |
Type of Fluid | Air and neutral gases |
Humidity Specifications | |
Measuring Range | From 5 to 95% RH |
Analogue Output | From 0 to 100% RH |
Accuracy | ±1.5% RH if 59 ≤ T ≤ 77°F (if 15 ≤ T ≤ 25°C) |
Drift Linked to Temperature | ±0.04 x (T-20) % RH if T < 59 or T > 77°F (if T < 15 or T > 25°C) |
Unit of Measurement | % RH |