Designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. A multitude of automation and labor-saving features are provided to ensure effective testing for a wide variety of requirements and test conditions. Select the most appropriate model for your applications.
Designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. A multitude of automation and labor-saving features are provided to ensure effective testing for a wide variety of requirements and test conditions. Select the most appropriate model for your applications.
Programable testing, full remote control, Automatic Insulation Withstanding Tester This electrical safety test instrument series is designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. A multitude of automation and labor-saving Features are provided to ensure effective testing for a wide variety of requirements and test conditions. Select the most appropriate model for your applications.
Features
Output voltage | 0.20 to 5.00 kV |
Voltage output method | PWM switching method |
Transformer capacity | 500 VA (rated 30 minutes) |
Output capacity | 50 VA (continuous) |
Voltage adjustment method | Digital setting (0.01 kV setting resolution) |
Output voltage accuracy | ±1.5% of setting voltage ±2 dgt. |
Voltage change rate | AC: ±7% or less(max. 5 kV at 100 mA →unloaded: with resistive load)*2 DC: ±16% or less(max. 5 kV at10 mA → unloaded: with resistive load)*2 |
Voltage waveform | Sine wave (5% or less distortion, unloaded) |
Voltage frequency | 50 or 60 Hz ±0.2% |
Output ripple voltage | 6% of output voltage or less (at 5 kV DC, 10 mA, resistive load) |
Output current | AC: 100 mA *1 DC:10 mA (continuous) |
Voltmeter | Digital: 0.00 to 5.00 kV (full scale) Accuracy: ±1.5% f.s. Analog: 0 to 5 kV (full scale) Accuracy: ±5% f.s. |
Current measurement range | AC: 0.01 to 100.0 mA DC: 0.01 to 10.0 mA |
Indicated value range | AC: 10 or 100 mA DC: 10 mA |
Measurement resolution | 0.00 to 10.00 or 0.01 mA (10-mA range) 10.1 to 100.0 or 0.1 mA (100-mA range, AC only) |
Current measurement accuracy | ± (2% rdg. + 5 dgt.) common to each range *3 |
Output voltage | 0.20 to 5.00 kV |
Voltage output method | PWM switching method |
Transformer capacity | 500 VA (rated 30 minutes) |
Output capacity | 50 VA (continuous) |
Voltage adjustment method | Digital setting (0.01 kV setting resolution) |
Output voltage accuracy | ±1.5% of setting voltage ±2 dgt. |
Voltage change rate | AC: ±7% or less(max. 5 kV at 100 mA →unloaded: with resistive load)*2 DC: ±16% or less(max. 5 kV at10 mA → unloaded: with resistive load)*2 |
Voltage waveform | Sine wave (5% or less distortion, unloaded) |
Voltage frequency | 50 or 60 Hz ±0.2% |
Output ripple voltage | 6% of output voltage or less (at 5 kV DC, 10 mA, resistive load) |
Output current | AC: 100 mA *1 DC:10 mA (continuous) |
Voltmeter | Digital: 0.00 to 5.00 kV (full scale) Accuracy: ±1.5% f.s. Analog: 0 to 5 kV (full scale) Accuracy: ±5% f.s. |
Current measurement range | AC: 0.01 to 100.0 mA DC: 0.01 to 10.0 mA |
Indicated value range | AC: 10 or 100 mA DC: 10 mA |
Measurement resolution | 0.00 to 10.00 or 0.01 mA (10-mA range) 10.1 to 100.0 or 0.1 mA (100-mA range, AC only) |
Current measurement accuracy | ± (2% rdg. + 5 dgt.) common to each range *3 |