Gauge pressure module with a range from 0 to 300 psi (0 to 20 Bar) and a 4X Burst Rating. This module features a reference uncertainty of ±0.0175%.
Gauge pressure module with a range from 0 to 300 psi (0 to 20 Bar) and a 4X Burst Rating. This module features a reference uncertainty of ±0.0175%.
Fluke 750P Series Pressure Modules are easy to use. To measure pressure, connect the pressure module to a pressure source or hand pump and then connect the pressure module cable to the calibrator. Apply pressure from the pressure source and it is displayed digitally on the calibrator. At the touch of a button, the pressure may be displayed in up to 11 different engineering units. When used with the 750 Series Documenting Process Calibrators, pressure readings can be date/time stamped and stored electronically for later retrieval. This saves time, eliminates errors, and supports F85compliance with quality standards and regulations.
Features
Applications
Parameter Range | 0 to 300 psi (0 to 20 bar) |
Burst Rating | 4X |
Hi Side Media | 316 SS |
Lo Side Media | NA |
Total Uncertainty 1 year (15-35℃) | ± 0.035 % |
Model | Parameter/Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) | Total Uncertainty 1-year1 | Total Uncertainty 6-month (15-35 °C) | Total Uncertainty 6-month1 |
---|---|---|---|---|---|---|---|---|---|
Differential | |||||||||
750P00 | 0 to 1 in H2O (0 to 2.5 mBar) | 30X | Dry Air | Dry Air | ± 0.15 % | ± 0.3 % | ± 0.35 % | ± 0.25 % | ± 0.30 % |
750P01 | 0 to 10 in H2O (0 to 25 mBar) | 3x | Dry Air | Dry Air | ± 0.1% | ± 0.2% | ± 0.3% | ± 0.15% | ± 0.25% |
750P02 | 0 to 1 psi (0 to 70 mBar) | 3x | Dry Air | Dry Air | ± 0.050% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750P22 | 0 to 1 psi (0 to 70 mBar) | 3x | 316 SS | Dry Air | ± 0.050% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750P03 | 0 to 5 psi (0 to 350 mBar | 3x | Dry Air | Dry Air | ± 0.02% | ± 0.04% | ± 0.05% | ± 0.035% | ± 0.04% |
750P23 | 0 to 5 psi (0 to 350 mBar) | 4x | 316 SS | Dry Air | ± 0.02% | ± 0.04% | ± 0.05% | ± 0.035% | ± 0.04% |
750P04 | 0 to 15 psi (0 to 1 bar) | 3x | Dry Air | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P24 | 0 to 15 psi (0 to 1 bar) | 4X | 316 SS | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Gage | |||||||||
750P05 | 0 to 30 psi (0 to 2 bar) | 4x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P06 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P08 | 0 to 1000 psi (0 to 70 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P09 | 0 to 1500 psi (0 to 100 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P2000 | 0 to 2000 psi (0 to 140 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
High | |||||||||
750P29 | 0 to 3000 psi (0 to 200 bar) | 3x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P30 | 0 to 5000 psi (0 to 340 bar) | 3x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P31 | 0 to 10000 psi (0 to 700 bar) | 2x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Absolute | |||||||||
750PA3 | 0 to 5 psia (0 to 350 mBar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA4 | 0 to 15 psia (0 to 1 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA5 | 0 to 30 psia (0 to 2 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA6 | 0 to 100 psia (0 to 7 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA27 | 0 to 300 psia (0 to 20 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA7 | 0 to 500 psia (0 to 35 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA8 | 0 to 1000 psia (0 to 70 bar) | 3X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA9 | 0 to 1500 psia (0 to 100 bar) | 3X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
Vacuum | |||||||||
750PV3 | -5 psi (-350 mBar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PV4 | -15 psi (-1 bar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
Dual | |||||||||
750PD2 | -1 to 1 psi (-70 to 70 mBar) | 4X | 316 SS | Dry Air | ± 0.05% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750PD3 | -5 to 5 psi (-350 to 350 mBar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PD10 | -10 to 10 psi (-700 to 700 mBar) | 4X | 316 SS | Dry Air | ± 0.025% | ± 0.05% | ± 0.07% | ± 0.04% | ± 0.06% |
750PD4 | -15 to 15 psi (-1 to 1 bar) | 4X | 316 SS | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD5 | -15 to 30 psi (-1 to 2 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD50 | -15 to 50 psi (-1 to 3.5 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD6 | -15 to 100 psi (-1 to 7 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD7 | -15 to 200 psi (-1 to 14 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD27 | -15 to 300 psi (-1 to 20 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Reference | |||||||||
750R045 | 0 to 15 psi (0 to 1 bar) | 3X | Dry Air | Dry Air | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R065 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R085 | 0 to 1000 psi (0 to 70 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R29 | 0 to 3000 psi (0 to 200 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R30 | 0 to 5000 psi (0 to 340 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS | |||
750R315 | 0 to 10000 psi (0 to 700 bar) | 2X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD5 | -15 to 30 psi (-1 to 2 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD65 | -12 to 100 psi (-1 to 7 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD27 | -12 to 300 psi (-0.8 to 20 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
Click on a category to view a selection of compatible accessories with the Fluke 750P27 Gauge Pressure Module, 0-20 Bar, 4X Burst Rating.
Parameter Range | 0 to 300 psi (0 to 20 bar) |
Burst Rating | 4X |
Hi Side Media | 316 SS |
Lo Side Media | NA |
Total Uncertainty 1 year (15-35℃) | ± 0.035 % |
Model | Parameter/Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) | Total Uncertainty 1-year1 | Total Uncertainty 6-month (15-35 °C) | Total Uncertainty 6-month1 |
---|---|---|---|---|---|---|---|---|---|
Differential | |||||||||
750P00 | 0 to 1 in H2O (0 to 2.5 mBar) | 30X | Dry Air | Dry Air | ± 0.15 % | ± 0.3 % | ± 0.35 % | ± 0.25 % | ± 0.30 % |
750P01 | 0 to 10 in H2O (0 to 25 mBar) | 3x | Dry Air | Dry Air | ± 0.1% | ± 0.2% | ± 0.3% | ± 0.15% | ± 0.25% |
750P02 | 0 to 1 psi (0 to 70 mBar) | 3x | Dry Air | Dry Air | ± 0.050% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750P22 | 0 to 1 psi (0 to 70 mBar) | 3x | 316 SS | Dry Air | ± 0.050% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750P03 | 0 to 5 psi (0 to 350 mBar | 3x | Dry Air | Dry Air | ± 0.02% | ± 0.04% | ± 0.05% | ± 0.035% | ± 0.04% |
750P23 | 0 to 5 psi (0 to 350 mBar) | 4x | 316 SS | Dry Air | ± 0.02% | ± 0.04% | ± 0.05% | ± 0.035% | ± 0.04% |
750P04 | 0 to 15 psi (0 to 1 bar) | 3x | Dry Air | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P24 | 0 to 15 psi (0 to 1 bar) | 4X | 316 SS | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Gage | |||||||||
750P05 | 0 to 30 psi (0 to 2 bar) | 4x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P06 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P08 | 0 to 1000 psi (0 to 70 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P09 | 0 to 1500 psi (0 to 100 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P2000 | 0 to 2000 psi (0 to 140 bar) | 4X | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
High | |||||||||
750P29 | 0 to 3000 psi (0 to 200 bar) | 3x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P30 | 0 to 5000 psi (0 to 340 bar) | 3x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750P31 | 0 to 10000 psi (0 to 700 bar) | 2x | 316 SS | N/A | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Absolute | |||||||||
750PA3 | 0 to 5 psia (0 to 350 mBar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA4 | 0 to 15 psia (0 to 1 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA5 | 0 to 30 psia (0 to 2 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA6 | 0 to 100 psia (0 to 7 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA27 | 0 to 300 psia (0 to 20 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA7 | 0 to 500 psia (0 to 35 bar) | 4X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA8 | 0 to 1000 psia (0 to 70 bar) | 3X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PA9 | 0 to 1500 psia (0 to 100 bar) | 3X | 316 SS | N/A | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
Vacuum | |||||||||
750PV3 | -5 psi (-350 mBar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PV4 | -15 psi (-1 bar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
Dual | |||||||||
750PD2 | -1 to 1 psi (-70 to 70 mBar) | 4X | 316 SS | Dry Air | ± 0.05% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
750PD3 | -5 to 5 psi (-350 to 350 mBar) | 4X | 316 SS | Dry Air | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
750PD10 | -10 to 10 psi (-700 to 700 mBar) | 4X | 316 SS | Dry Air | ± 0.025% | ± 0.05% | ± 0.07% | ± 0.04% | ± 0.06% |
750PD4 | -15 to 15 psi (-1 to 1 bar) | 4X | 316 SS | Dry Air | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD5 | -15 to 30 psi (-1 to 2 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD50 | -15 to 50 psi (-1 to 3.5 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD6 | -15 to 100 psi (-1 to 7 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD7 | -15 to 200 psi (-1 to 14 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
750PD27 | -15 to 300 psi (-1 to 20 bar) | 4X | 316 SS | - | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
Reference | |||||||||
750R045 | 0 to 15 psi (0 to 1 bar) | 3X | Dry Air | Dry Air | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R065 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R085 | 0 to 1000 psi (0 to 70 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R29 | 0 to 3000 psi (0 to 200 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750R30 | 0 to 5000 psi (0 to 340 bar) | 3X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS | |||
750R315 | 0 to 10000 psi (0 to 700 bar) | 2X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD5 | -15 to 30 psi (-1 to 2 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD65 | -12 to 100 psi (-1 to 7 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
750RD27 | -12 to 300 psi (-0.8 to 20 bar) | 4X | 316 SS | - | ±0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
Click on a category to view a selection of compatible accessories with the Fluke 750P27 Gauge Pressure Module, 0-20 Bar, 4X Burst Rating.